Medical Device 'Significant Changes' – Navigating EU MDR Article 120(3) using MDCG 2020-3 rev. 1

Shen May Khoo
A pile of question marks

With the transition to the EU MDR, legacy devices with certificates issued under the old EU directives (Directive 93/42/EEC on medical devices, MDD, and Directive 90/385/EEC on active implantable medical devices, AIMDD) can stay on the market under the transitional provisions set out in Article 120(3) of the MDR, provided they do not undergo any significant changes.

In this article, we explain the concept of a ‘significant change’ and, using examples, help you understand whether you can still take advantage of the transitional arrangements in MDR Article 120(3).

Transition periods

The transition periods in Article 120(3) MDR as per Regulation (EU) 2023/607 are as follows:

  • Class III and IIb implants: new transition period until 31 December 2027
  • Class IIb and lower: new transition period until 31 December 2028
  • Class III custom-made products: new transition period until 31 December 2026
  • No ‘sell-off’ deadline

What does significant change mean?

A significant change in the design or intended purpose consists of two cumulative elements:

  • There is a change in the design or intended purpose, and
  • The change is significant

Therefore, changes that do not concern the design or intended purpose are out of scope of Article 120(3) of the MDR. Changes that concern the design or intended purpose only fall under Article 120(3) if they are considered ‘significant’.

Responsibilities of medical device manufacturers regarding ‘significance’ of changes

Medical device manufacturers have two primary responsibilities regarding changes:

  • provide evidence and justification that a change does not affect the design or intended purpose, or
  • in cases where the change affects the design or intended purpose of the device, that it is non-significant.

The outcome of this assessment should be documented and made available to a competent authority when requested.

If a change to a legacy device is not a significant change in design or intended purpose, it can be implemented using the above transitional arrangement. In these instances, manufacturers must adhere to the documentation criteria of AIMDD/MDD, meaning that the revised technical documentation must enable the evaluation of the product’s compliance with the relevant standards.

Determining a significant change to the design or intended purpose

The MDCG 2020-3 rev. 1 guidance document sets out whether a change in the design or intended purpose of a device is considered ‘significant’ within the meaning of EU MDR Article 120(3c), point (b).

Examples of changes in design and/or intended purpose that are ‘non-significant’:

  • Changes related to corrective actions assessed and accepted by the competent authority
  • Correction of spelling mistakes or merely editorial changes of the information to be supplied with the device (e.g. label or instructions for use)
  • Clarifications of intended purpose, population, or clinical application in the information to be supplied with the device in line with the original certification
  • Updates of the information to be supplied with the device (e.g. label, instructions for use or implant card) if they are required by EU legislation other than the MDR, are mere clarifications and do not adversely affect the devices’ safety and performance in relation to existing or new risks

The following diagram, taken from the MDCG 2020-3 rev. 1, helps understand when changes in design and/or intended purpose may be considered ‘significant’:

Design changes and changes of the intended purpose which may be considered 'significant' when interpreting Art. 120(3c), point (b) MDR – Main Chart
Design changes and changes of the intended purpose which may be considered 'significant' when interpreting Art. 120(3c), point (b) MDR – Main Chart taken from MDCG 2020-3 rev. 1

Charts A to E in the MDCG guideline contain specific areas where changes in the intended purpose, device design, device software, changes related to a substance or material or changes related to sterilisation are likely to be held to be significant. It’s vital to consult these charts when understanding whether a change will be significant and, by extension, whether your device can make use of the transitional arrangement.

In summary

The definition of a “significant change” involves two cumulative elements: a change in design or intended purpose that is deemed significant. Manufacturers bear the responsibility of determining and justifying whether a change in their medical device is significant or not, and therefore whether relevant technical documents require updating in line with the changes. The MDCG 2020-3 rev. 1 guidance document provides guidance on determining the significance of changes in design or intended purpose.

For further information regarding how we can meet your MDR requirements, including advice on whether changes are likely to be significant, please contact us for a free and confidential discussion.

Related articles

  1. A man carefully steps across a cliff-face. An analogy for assessing risk.

    Top 5 Common Pitfalls to Avoid During Risk Assessment

    Learn how to sidestep costly mistakes which manufacturers commonly make. From hazard ID to post-market surveillance, we help you improve safety and speed up approvals.

    Kamiya Crabtree Kamiya Crabtree Regulatory Medical Writer
  2. A label maker printing bar-code labels.

    Labelling 101: A Comprehensive Overview for Medical Device Manufacturers

    Labelling and packaging are critical elements to ensuring safety, compliance, and ease of use.

    Kamiya Crabtree Kamiya Crabtree Regulatory Medical Writer
  3. An illustration of a brain-shaped object on an abstract background.

    European Commission Guidelines on Prohibited Artificial Intelligence Practices

    Summary of the 8 AI practices prohibited by the EU 2024/1689 artificial intelligence (AI) Act.

    Dr Clare Dixon Dr Clare Dixon Regulatory Specialist
  4. A compass being used to navigate across mountainous countryside.

    Navigating Non-Conformities in Technical Documentation

    We explore how to manage non-conformities effectively and implement Corrective and Preventive Actions (CAPAs).

    Kamiya Crabtree Kamiya Crabtree Regulatory Medical Writer
  5. A photograph of a literal maze that we're using as a clever metaphor.

    Mastering the EU MDR: Essential Steps for Compliance-Ready Docs

    If you're uncertain about the readiness of your EU MDR documentation, this article provides an overview of the essential steps to ensure you’re on track.

    Kamiya Crabtree Kamiya Crabtree Regulatory Medical Writer
  6. An illustration showing scientists at work.

    A Guide to Electronic Instructions for Use (eIFU)

    Electronic Instructions for Use (eIFUs) are set to revolutionise how medical device instructions are delivered. We explore what this means for you.

    Dr Will Brambley Dr Will Brambley Lead Medical Writer
  7. Two helicopters look as if they are about to collide: An analogy for risk.

    Navigating Risk Management Requirements under the EU MDR

    This is a cornerstone of EU MDR 2017/745, requiring a continuous, well-documented approach. We unpack key requirements and provide actionable strategies.

    Peter Boxall Peter Boxall Lead Medical Writer
  8. A doctor operates a tablet computer.

    Beyond the Acronyms: Understanding SaMD and SiMD

    As software advancements continue, the line between traditional hardware-centric medical devices and software-driven solutions becomes increasingly blurred.

    Kamiya Crabtree Kamiya Crabtree Regulatory Medical Writer
  9. A team of profesional-looking people sit around a table, congratulating themselves.

    Extending the Validity of your IVDD Certificates – Key Dates

    The EU and the MHRA have extended the validity of IVDD certificates, allowing you more time to transition to the IVDR. We explain what this means for manufacturers.

    Kamiya Crabtree Kamiya Crabtree Regulatory Medical Writer
  10. A team of profesional-looking people sit around a table, congratulating themselves.

    GSPR 1: A New Era of Performance with Safety at the Core

    This regulation emphasizes risk management, durable design & biocompatibility to ensure medical devices are safe and effective. GSPR 1 protects users while driving innovation in medical technology.

    Kamiya Crabtree Kamiya Crabtree Regulatory Medical Writer
  11. Cybersecurity Vulnerabilities in Medical Devices: FDA Alerts on Contec and Epsimed Monitors

    Patients can be exposed to risks when devices are online. We explore implications for EU MDR/IVDR cybersecurity requirements, including MDCG guidance

    Dr Clare Dixon Dr Clare Dixon Regulatory Specialist
  12. A futuristic-looking factory full of labelled cardboard boxes.

    Decoding UDI: Your Ultimate Guide to Smarter Medical Device Labelling

    The Unique Device Identifier (UDI) ensures medical device traceability and compliance. We break down its structure, Device Identifier (UDI-DI), Production Identifier (UDI-PI) and its role in EUDAMED.

    Kamiya Crabtree Kamiya Crabtree Regulatory Medical Writer
  13. A medical team discuss performance data at their desktop computer.

    Key Updates for Navigating EMDN: MDCG 2024-2 Rev.1 & 2021-12 Rev.1

    Release of the updated guidance helps manufacturers navigate the EMDN system for accurate device classification, ensuring market access.

    Ron Sangal Ron Sangal Lead Medical Writer
  14. A dated monitor for medical equipment.

    Understanding Clinical Evidence Requirements with MDCG 2020-6

    How can manufacturers ensure legacy devices meet MDR's stringent requirements? Discover how MDCG 2020-6 guidance simplifies the path to compliance.

    Dr Clare Dixon Dr Clare Dixon Regulatory Specialist
  15. A stethoscope laid on a desk of regulatory documentation.

    Clinical benefits of an in vitro diagnostic medical device

    How to determine the clinical benefit of an IVD and successfully incorporate it into regulatory documentation.

    Dr Gayle Buchel Dr Gayle Buchel Chief Medical Writer
  16. EU flags

    Regulation (EU) 2024/1860 - Its impact on EU MDR and IVDR

    How does the recent Regulation (EU) 2024/1860 amendment affect the EU MDR & IVDR?

    Shona Richardson Shona Richardson Regulatory Medical Writer
  17. EU flag

    MDCG 2024-10 - Orphan medical devices

    How to apply MDR pre-market clinical evidence requirements to medical devices intended for limited usage.

    Dr Simon Cumiskey Dr Simon Cumiskey Senior Lead Medical Writer
  18. Considering a medical device's intended purpose

    A medical device's intended purpose - what is the point?

    How do you define intended purpose, indication for use, intended clinical benefits, and claims?

    Dr Simon Cumiskey Dr Simon Cumiskey Senior Lead Medical Writer
  19. Mantra Systems presents EnableChat, your AI-powered MDR & MDCG chatbot

    EnableChat - Your AI-powered MDR and MDCG chatbot

    Search the MDR and MDCG documents in seconds by asking EnableChat your questions.

    Dr Simon Cumiskey Dr Simon Cumiskey Senior Lead Medical Writer
  20. Searching adverse event databases for vigilance data

    Staying vigilant - A guide to searching for adverse events data

    We discuss the pros and cons of existing adverse event databases for vigilance data searching.

    Dr Simon Cumiskey Dr Simon Cumiskey Senior Lead Medical Writer
  21. A doctor reading an SSCP document with a patient

    What is Summary of Safety and Clinical Performance (SSCP)?

    We explain what the SSCP is, when you'll need it and what its objectives are.

    Sandra Gopinath Sandra Gopinath Chief Regulatory Officer
  22. A signpost giving unsure directions

    MDR or IVDR - A sibling rivalry?

    A guide to easily understanding whether your device is a medical device or an in vitro diagnostic medical device (IVD).

    Dr Gayle Buchel Dr Gayle Buchel Chief Medical Writer
  23. An EU and UK flag

    What the latest Brexit U-turn means for CE Marking of medical devices in Great Britain

    Will Great Britain continue to allow the use of the CE mark for medical devices beyond the 2024 deadline?

    Dr Hanna Gul Dr Hanna Gul Lead Medical Writer
  24. A woman writing her own medical device regulation documentation

    Gain confidence, reassurance and control over your EU MDR strategy

    Find out how to build your own technical files within a guided framework while minimising financial outlays.

    Dr Gayle Buchel Dr Gayle Buchel Chief Medical Writer
  25. Racing to achieve MDR compliance

    Still racing to achieve MDR compliance? A transition period update

    On January 6th 2023, the EU commission has adopted the proposal to extend the transition rules of the EU MDR.

    Sandra Gopinath Sandra Gopinath Chief Regulatory Officer
  26. A 7-step guide to navigating regulatory requirements for medical device start-ups

    A medical device regulations guide for start-up companies

    We present a 7-step guide to navigating regulatory requirements on a budget.

    Dr Paul Hercock Dr Paul Hercock Chief Executive Officer
  27. An update on UKCA Marking of Medical Devices

    UKCA Marking of Medical Devices – An update on the status quo

    We review recently updated requirements for UKCA marking and what it means for your regulatory strategy.

    Dr Hanna Gul Dr Hanna Gul Lead Medical Writer
  28. How to choose a CER writer for your MDR Clinical Evaluation

    Choosing a CER writer for your MDR Clinical Evaluations

    We've compiled a list of considerations that will help you make the right choice when choosing a CER writer.

    Dr Paul Hercock Dr Paul Hercock Chief Executive Officer
  29. Achieving MDR Compliance for Class I medical devices

    How to achieve MDR Compliance for Class I medical devices

    We outline a strategy for the regulatory compliance of Class I medical devices.

    Sandra Gopinath Sandra Gopinath Chief Regulatory Officer
  30. Literature Search, SOTA Review and Clinical Evaluation

    Literature Search, SOTA Review process and Clinical Evaluation

    We help to demystify the process of systematic search & review of literature for Clinical Evaluation.

    Sandra Gopinath Sandra Gopinath Chief Regulatory Officer
  31. Literature Search Protocols & SOTA Reviews for medical devices and what to know before you start

    Literature searches and reviews for medical devices - what to know before you start

    We explain what you should know before beginning a literature search & review for your medical device.

    Sandra Gopinath Sandra Gopinath Chief Regulatory Officer
  32. Five useful resources when writing a medical device CER

    Five useful resources when writing a medical device CER

    We outline five of the most useful and trustworthy Clinical Evaluation Report writing resources.

    Dr Victoria Cartwright Dr Victoria Cartwright Relationship Manager
  33. Avoid pitfalls when writing a Clinical Evaluation Report

    Five common pitfalls when writing a Clinical Evaluation Report

    We illustrate five pitfalls when writing CERs and give you some tips to overcome them.

    Dr Paul Hercock Dr Paul Hercock Chief Executive Officer
  34. How to make a medical device equivalence claim under the MDR

    Five tips for making a medical device equivalence claim under the MDR

    We'll show you what to keep in mind with regards to equivalance and Clinical Evaluation.

    Sandra Gopinath Sandra Gopinath Chief Regulatory Officer
  35. Keeping medical devices in market and maintaining CE-marks - a guide to effective data collection

    Keeping medical devices in market and maintaining CE-marks

    The 4 golden rules to drive regulatory compliance with PMCF and vigilance data collection.

    Dr Paul Hercock Dr Paul Hercock Chief Executive Officer
  36. How PMCF goes beyond simple compliance - improving products and engaging customers

    How PMCF goes beyond simple compliance

    The wider benefits of a well-designed PMCF system include improving your products and your relationship with your clients.

    Dr Paul Hercock Dr Paul Hercock Chief Executive Officer
  37. PMCF systems for medical devices

    Why you'll almost certainly need a PMCF system for your medical devices

    We tell you what to be aware of under the EU MDR regarding PMCF and your medical devices.

    Dr Paul Hercock Dr Paul Hercock Chief Executive Officer
  38. Ensure medical device regulatory compliance of your devices through Brexit

    The impact of Brexit on medical device regulatory compliance

    How to ensure regulatory alignment of your devices in the territories affected by Brexit.

    Dr Paul Hercock Dr Paul Hercock Chief Executive Officer
  39. Use medical device regulatory consulting services to supercharge your MDR transition

    Is outside consulting support the answer to your MDR transition?

    Getting ready for the MDR is a demanding process. Outsourcing might be your solution.

    Dr Paul Hercock Dr Paul Hercock Chief Executive Officer
  40. Increasing data entry compliance in PMCF studies

    Increasing data entry compliance in PMCF studies

    5 methods every medical device manufacturer should know to improve their Post-Market Clinical Follow-up studies.

    Dr Paul Hercock Dr Paul Hercock Chief Executive Officer
  41. Why medical doctors can drive MDR compliance

    Why medical doctors can drive MDR compliance

    Working with the MDR requires knowing how to work with clinical evidence. Medical doctors are perfectly positioned to meet this requirement.

    Dr Victoria Cartwright Dr Victoria Cartwright Relationship Manager
  42. Software as a Medical Device

    Software as a Medical Device

    Unless you have spent time working with medical device legislation in the past, the idea that software could be a medical device may be rather unexpected.

    Dr Paul Hercock Dr Paul Hercock Chief Executive Officer
  43. clinical investigator for pmcf eu mdr compliance

    Ensuring that clinical investigations work in practice

    How can medical device manufacturers ensure valid clinical investigations when access to medical expertise remains limited?

    Dr Paul Hercock Dr Paul Hercock Chief Executive Officer
  44. Coronavirus and medical device regulations

    Relaxing medical device regulatory requirements during a healthcare crisis

    During the coronavirus pandemic, how far should we go when relaxing medical device regulatory requirements?

    Dr Paul Hercock Dr Paul Hercock Chief Executive Officer
  45. The new MDR compliance challenge

    The new MDR compliance challenge

    Across the industry, medical device companies are facing challenges in meeting the demands of the new Medical Device Regulations (MDR) 2017/745 framework.

    Dr Paul Hercock Dr Paul Hercock Chief Executive Officer
  46. Sources of Real World Evidence for MDR compliance

    Sources of Real World Evidence for MDR compliance

    At Mantra Systems our objective is to make sure that our clients choose the method of real world data harvesting that is right for them.

    Dr Paul Hercock Dr Paul Hercock Chief Executive Officer

More articles

Do you need support with your medical device approval strategy?

Contact us today